Overview:

Measuring how materials emit light over time and across different wavelengths reveals the hidden processes happening inside them. By using extremely short pulses of light, time-resolved photoluminescence captures how quickly this emission fades, giving precise insight into the material’s kinetics. The choice of detection method depends on how fast the glow decays, and time-correlated single-photon counting (TCSPC)stands out as a highly sensitive technique capable of detecting even the tiniest and fastest changes in light emission. This approach allows to peek into the dynamic behaviour of materials at the molecular or electronic level, making invisible processes visible.

On the other hand Two-photon photoluminescence (2-photon PL) uses two low-energy photons that arrive at the same time to excite a material. This allows to probe deeper into the material with minimal photodamage, as two-photon excitation occurs only at the focal point where the photon density is highest.

TCSPC and 2-photon PL are both powerful ways to study howmaterials interact with light, but they reveal different kinds of information.

Description: 

A Spectra Physics Mai Tai HP outputs 100 fs pulses at 80MHz, with up to 2.75 W of average power, tuneable across 690-1040 nm. The Mai Tai output is directed to one of two home-built experimental systems for optical materials characterisation - TCSPC, and 2-photon PL;

Time-correlated single-photon counting, or TCSPC, is a technique which records the time delay between photoexcitation of a sample with a pulsed laser source, and the subsequent emission of light as photoluminescence. Repeated at high repetition rate, a histogram of emission times provides a temporal emission profile of the material, from which the lifetime of photoexcited states in the sample can be extracted. For this technique, the Mai Tai output is passed through a pulse picker, for repetition rate reduction, and then a second-harmonic generation unit for frequency conversion for 345 - 520 nm excitation of samples. a monochromator and MCP-PMT enable recording of the resultant photoluminescnece with an IRF of approximately 50 ps.

2-photon PL is a technique where an ultrashort laser pulse is focussed onto a sample such that the excitation density is high enough to cause simultaneous absorption of two excitation photons. The subsequent photoluminescence is then recorded. For this setup the Mai Tai output is directed into a microscope objective to focus the light onto the sample. the subsequent PL is collected collinearly through the objective, separated by a dichroic mirror, and directed to an Ocean Optics QE-Pro spectrometer to record the 2-photon PL emission spectrum.

Capabilities/ key specifications:
•High Temporal Resolution – Can measure fluorescence or photoluminescence decay on the picosecond (10⁻¹² s) scale, capturing ultrafast processes.
•Extreme Sensitivity – Detects single photons, allowing measurement of very weak emissions that other techniques might miss.
•Accurate Kinetic Information – Provides true decay kinetics, making it possible to study fast dynamic processes in molecules and materials.
•Low Sample Damage – Requires very low light intensity, reducing the risk of photobleaching or altering the sample.
•Sample Flexibility – Works for thin films, solutions, and other sample types.
•High Throughput – Capable of fast data acquisition, making measurements efficient.

Uses/ applications:
•Fluorescence Lifetime Measurements – Determining lifetimes of molecules, proteins, dyes, and quantum dots and investigating excited-state dynamics and energy transfer pathways.
•Material Science – Studying photoluminescence dynamics in semiconductors, thin films, and nanomaterials.
•Chemical Analysis – Tracking fast photochemical reactions and transient species.
•Optoelectronics – Evaluating LEDs, solar cells, and other light-emitting devices.
•Single-Molecule Detection – Observing ultra-weak emissions at the single-photon level.
•Radioactive Materials – Studying radioluminescence and decay processes in radioactive samples.

Instrument(s) make and model:
Spectra-Physics Mai Tai High Performance (HP) mode-locked Ti:Sapphire femtosecond laser: Pulse Width <100 fs, 720 – 1040 nm at 80 MHz repetition rate, Average Power at 800 nm 2.5 W.
A.P.E PulseSelect: Pulse picker uses an acoustic-optic modulator, wavelength range 500nm-1600nm, Max. input power 5W at 800 nm, Input frequency 72MHz-81MHz, Output Rep rate 308 Hz -40 MHz (based on 80MHz input frequency).
APE Harmonic Generator: Second Harmonics Generator, input wavelength range 600 nm-2000nm, Output wavelength range 330nm - 1000nm, Pulse width - fs or ps.
SPEX Spectrometer: Focal Length of 0.25 meters, wavelength Range from 200 nm to 800 nm, depending on the grating used, Resolution: Varies with grating and slit width; typically, in the range of 0.1 nm to 1 nm.
Hamamatsu R3809U-5 MCP-PMT: A micro-channel plate photo-multiplier tube with a detection range of 160 – 850 nm.
Ocean Optics QE-pro: 190-1100nm, FWHM 1.2-6.87 nm, int time 8ms -60min.

Facility keywords

  • TCSPC
  • lifetime
  • photoluminescence
  • 2-photon

Research Highlights, Institutes and Platforms

  • Photon Science Institute

Fingerprint

Explore the research areas in which this equipment has been used. These labels are generated based on the related outputs. Together they form a unique fingerprint.