Through the use of specialised spin-coated polymer coatings, masks for evaporation and etching can be written onto user’s samples using an electron beam (e-beam). Firstly, a design is made in one of many commercially available pieces of design software and converted to the correct file-type for the machine; once loaded, the machine will fracture the design into small areas that it will sequentially scan and, therefore, expose with the e-beam. The pattern is then developed in a wet chemical developer solution. ‘Positive’ e-beam resist coatings become more soluble in a developer solution when exposed to an e-beam, ideal for patterning electrodes and other structures onto devices which can later be metallised using one of a multitude of metal deposition techniques available; ‘negative’ e-beam resist coatings become less soluble in a developer solution when exposed to an e-beam, generally more useful when a small area needs to be protected during wet/dry etch processes, or when the majority of a sample requires metallization. Once metallization/etching is complete, the remaining resist can be removed with a remover solution, this will also remove any metal remaining on top of the resist, a process known as lift-off. The Sigma has the capability to provide acceleration voltages up to 30kV, but is generally used at 10kV by most of the current users as, although it reduces the ultimate resolution of the system through more backscattered electrons in the resist and a larger forward scattering angle, it does improve the contrast of alignment markers needed to properly align designs to the substrate and gives improved lift-off. This machine is designed to pattern smaller device areas (up to a maximum of 2mmx2mm at a time), larger areas (up to 8” wafers) should be patterned on the Raith EBPB5200. The Sigma machine is very heavily used in the fabrication of the 2D devices the NGI users produce, it is ideal for the fabrication of devices not requiring the large area, speed or ultimate resolution offered by the larger Raith machine.
The Sigma has a Gemini I e-beam column and is equipped with an Oxford Instruments Energy Dispersive X-ray Spectrometer (EDS) for elemental analysis. The Sigma also includes a chamber secondary electrons detector (SED), an in-lens secondary electrons detector and a backscattered electrons detector.
Electron Source: Field Emission Source
Acceleration Voltage: 0.2 – 30 kV
Beam Current: 3 pA – 20 nA
Maximum Field Size: 2 mm x 2 mm
Stage Travel Range: 125 mm x 125 mm
User Interface: Zeiss SmartSEM + Raith ELPHY Quantum NANOSUITE
Thermal Stability: ~0.2%/h
Minimum Feature Size (Lithography): ~50-100 nm
Minimum Feature Size (Imaging): ~1 nm @ 30kV
Stitching Accuracy: ~1 µm
Maximum Substrate Size: 50 mm x 50 mm
The Sigma has a Gemini I e-beam column and is equipped with an Oxford Instruments Energy Dispersive X-ray Spectrometer (EDS) for elemental analysis. The Sigma also includes a chamber secondary electrons detector (SED), an in-lens secondary electrons detector and a backscattered electrons detector.
Electron Source: Field Emission Source
Acceleration Voltage: 0.2 – 30 kV
Beam Current: 3 pA – 20 nA
Maximum Field Size: 2 mm x 2 mm
Stage Travel Range: 125 mm x 125 mm
User Interface: Zeiss SmartSEM + Raith ELPHY Quantum NANOSUITE
Thermal Stability: ~0.2%/h
Minimum Feature Size (Lithography): ~50-100 nm
Minimum Feature Size (Imaging): ~1 nm @ 30kV
Stitching Accuracy: ~1 µm
Maximum Substrate Size: 50 mm x 50 mm