A multi-centre dosimetry audit on advanced radiotherapy in lung as part of the Isotoxic IMRT study

Corinne Faivre-Finn, Y Tsang, Antony Carver, Nicki Groom, Catherine Harris, D Eaton

Research output: Contribution to journalArticlepeer-review

Abstract

Background and purpose: In lung cancer patients, the accuracy of dose calculation by radiotherapy treatment planning systems may be reduced by the presence of tissue inhomogeneities. Intensity-Modulated Radiotherapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT) commonly include the use of small field segments which can exacerbate these issues. A multicentre dosimetry audit was undertaken to verify the accuracy of treatment delivery at participating centres.

Materials and methods: An inhomogeneous phantom with regions of lung equivalent materials was used. The audit consisted of two parts: basic beam model tests and clinical trial plan measurements. An ArcCheck diode array was used to measure fluences from the treatment delivery. Ion chamber and film planar dose readings near the interface were acquired.

Results: Nine centres were visited, with a total of eleven distinct combinations of planning and delivery system. Significant differences were found in ion chamber dose measurements in basic beam model tests between treatment planning system algorithms. For the clinical trial plan test, gamma analysis of the entry and exit dose fluence showed good agreement, with mean pass rates of 97.2% (95% CI: 95.3–99.1) and 99.0% (95% CI: 98.3–99.7) for tolerances of 3%/2 mm and 3%/3 mm respectively. No significant differences were found between treatment planning system algorithms, delivery techniques and linac manufacturers.

Conclusion: This multi-centre dosimetry audit of complex IMRT/VMAT delivery provides confidence in the accuracy of modern planning and delivery systems in inhomogeneous tissues. The findings from this study can be used as a reference for future dosimetry audits.
Original languageEnglish
Pages (from-to)17-21
JournalPhysics and Imaging in Radiation Oncology
Volume4
Early online date24 Nov 2017
DOIs
Publication statusPublished - 24 Nov 2017

Research Beacons, Institutes and Platforms

  • Manchester Cancer Research Centre

Fingerprint

Dive into the research topics of 'A multi-centre dosimetry audit on advanced radiotherapy in lung as part of the Isotoxic IMRT study'. Together they form a unique fingerprint.

Cite this