Abstract
A system for quality assessment (QA) of the CoaguChek (Roche Diagnostics, Mannheim, Germany) point-of-care testing prothrombin time monitor has been developed by the European Concerted Action on Anticoagulation. Hitherto there has not been an adequate rapid method for CoaguChek QA. Sets of 5 certified international normalized ratio (INR) plasma samples were tested on 539 CoaguChek monitors by experienced staff at 9 Netherlands Thrombosis Centers and results compared with certified INR. A 15% or more deviation has been classified as significant deviation. Overall mean and certified INR values were similar, but 20.3% of participants showed a 15% or more deviation from the certified INR on at least 1 of the 5 QA plasma samples. Statistically significant differences in results with different lots of CoaguChek test strips were found. There is need for large scale QA of CoaguChek monitors. The importance of the 5 CoaguChek certified INR QA plasma samples being tested on a single occasion is demonstrated. © American Society for Clinical Pathology.
Original language | English |
---|---|
Pages (from-to) | 756-761 |
Number of pages | 5 |
Journal | American journal of clinical pathology |
Volume | 126 |
Issue number | 5 |
DOIs | |
Publication status | Published - Nov 2006 |