Analyzing and correcting spectrometer temperature sensitivity

Paul J. Bowyer, Alistair G. Swanson, Gareth A. Morris

    Research output: Contribution to journalArticlepeer-review

    Abstract

    Stability and reproducibility of the spectrometer are fundamental to the success of many modern NMR experiments. Variation in room temperature is a particularly important source of instability, in part because it can cause coherent artifacts in NMR spectra. Small changes in room temperature lead to corresponding changes in the phases, amplitudes, and frequencies of NMR signals. These can lead in turn to apparently random spectral artifacts such as t1-noise in two-dimensional (2D) NMR and to the incomplete cancelation of signals in difference spectra, but also, importantly, to F1 satellite signals in 2D spectra. These "parallel diagonals" arise from the use of air conditioning, which typically forces room temperature to oscillate within a fixed band. Work to identify, quantify, and suppress sources of temperature sensitivity in a modern 300-MHz spectrometer has led to a greater than 10-fold improvement in the signal-to-artifact ratio. © 2001 Academic Press.
    Original languageEnglish
    Pages (from-to)234-246
    Number of pages12
    JournalJOURNAL OF MAGNETIC RESONANCE
    Volume152
    Issue number2
    DOIs
    Publication statusPublished - 2001

    Keywords

    • Air Conditioning
    • Artifacts
    • Calibration
    • Equipment Design
    • instrumentation: Magnetic Resonance Spectroscopy
    • Reproducibility of Results
    • Temperature

    Fingerprint

    Dive into the research topics of 'Analyzing and correcting spectrometer temperature sensitivity'. Together they form a unique fingerprint.