Correcting for Frequency Drift in Clinical Brain MR Spectroscopy

Benjamin C Rowland, Huijun Liao, Fatah Adan, Laura Mariano, John Irvine, Alexander P Lin

Research output: Contribution to journalArticlepeer-review

Abstract

PURPOSE: Averaging multiple repetitions to improve signal-to-noise ratio is common practice in magnetic resonance spectroscopy (MRS). However, temporal variations in scanner B0 due to motion or gradient heating may cause spectra to become misaligned, broadening and distorting peaks and impacting on processing and quantification. We present a comparison using in vivo data of different methods for correcting these errors.

METHODS: Three different correction methods were applied to 53 brain scans: residual water peak alignment, creatine fitting, and spectral registration. In 32 of 53 subjects, diffusion tensor imaging (DTI) was acquired prior to the MRS scan. We compared the resulting linewidths to find the most effective technique. In addition, the impact on metabolite concentration estimates was evaluated.

RESULTS: MRS data acquired after DTI imaging exhibited a frequency drift four times higher than data without DTI, resulting in changes to metabolite concentrations, particularly glutamate/glutamine. All three correction methods produced significantly improved linewidths relative to uncorrected data, with spectral registration performing best by a small margin.

CONCLUSION: Frequency correction is an important step in processing MRS data, significantly impacting metabolite quantification, particularly after echo-planar imaging that often occurs with MRS scans in clinical studies. Spectral registration proved most effective at frequency correction.

Original languageEnglish
Pages (from-to)23-28
Number of pages6
JournalJournal of Neuroimaging
Volume27
Issue number1
Early online date7 Sept 2016
DOIs
Publication statusPublished - Jan 2017

Keywords

  • Algorithms
  • Brain
  • Diffusion Tensor Imaging
  • Humans
  • Magnetic Fields
  • Magnetic Resonance Spectroscopy
  • Reproducibility of Results
  • Signal-To-Noise Ratio
  • Journal Article
  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

Research Beacons, Institutes and Platforms

  • Manchester Cancer Research Centre

Fingerprint

Dive into the research topics of 'Correcting for Frequency Drift in Clinical Brain MR Spectroscopy'. Together they form a unique fingerprint.

Cite this