European concerted action on anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration

Leon Poller, Michelle Keown, Nikhil Chauhan, Anton M H P Van Den Besselaar, Armando Tripodi, Caroline Shiach, Jorgen Jespersen

    Research output: Contribution to journalArticlepeer-review

    Abstract

    The international normalized ratio (INR) on two widely used point-of-care test (POCT) prothrombin time (PT) monitors (CoaguChek Mini and TAS PT-NC) differed considerably and also differed from the 'true' INR obtained on the same samples using a manual PT and the same species thromboplastin international reference preparation. Agreement between the displayed INR and difference from 'true' INR has been reassessed following an independent international sensitivity index (ISI) calibration of the two systems. The displayed INRs taken at seven centres were compared with 'true' INRs from the same blood donations and INRs based on the resulting ISI. The overall difference between the displayed INRs on the two monitor systems was reduced from 21.0% to 3.5%. The overall difference in mean INR of system A from the 'true' INR was reduced from 19.0% to 9.5% and of system B from 6.8% to 0.3%, but individual centre's results still showed considerable mean INR variability. Differences between overall displayed INR with the two monitor systems have been reduced by an independent multicentre calibration, and agreement with 'true' INR on the same blood samples improved. However, marked variability in mean INR at individual centres remained after ISI correction, which demonstrates the need for external quality control of individual POCT whole-blood PT monitors.
    Original languageEnglish
    Pages (from-to)944-949
    Number of pages5
    JournalBritish Journal of Haematology
    Volume122
    Issue number6
    DOIs
    Publication statusPublished - Sept 2003

    Keywords

    • International normalized ratio
    • Multicentre international sensitivity index (ISI)
    • Point-of-care test (POCT) monitors
    • Prothrombin time

    Fingerprint

    Dive into the research topics of 'European concerted action on anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration'. Together they form a unique fingerprint.