Abstract
Background: The International Sensitivity Index (ISI) calibration of whole-blood prothrombin time (PT) monitors for point-of-care testing (POCT) described by Tripodi et al. (Thromb Haemost 1993;70:921-4) has been shown to be dependable but is too complex and demanding. The use of plasma would simplify calibration of whole-blood POCT PT monitors, but important differences may exist between the ISI for whole blood and plasma calibrations. Methods: In a 10-center calibration study of two POCT whole-blood monitoring systems (CoaguChek Mini and TAS PT-NC), we characterized the relationship between the log PT for whole blood and fresh plasma with use of single lots of test strips/cards. This relationship (linear) was used to correct the difference between the whole-blood and plasma ISI. The reliability of the correction with different lots of test strips/cards was assessed at three centers. The linear relationship was used to correct the difference in the whole-blood and plasma ISI with four other lots of TAS PT-NC cards and with two additional lots of CoaguChek Mini test strips. Results: The correction decreased the ISI difference from 13.3% to 0.9% for the TAS PT-NC and from 5.7% to 0.6% for the CoaguChek Mini. In assessments at three centers, which included different lots of test strips/cards, the mean ISI difference was markedly decreased with the TAS PT-NC but not with the CoaguChek Mini, for which the mean ISI difference increased slightly. Conclusions: The proposed correction resolves the discrepancy between whole-blood and fresh plasma ISI calibrations with TAS PT-NC test cards. The CoaguChek Mini systems could be calibrated without this correction. © 2002 American Association for Clinical Chemistry.
| Original language | English |
|---|---|
| Pages (from-to) | 1672-1680 |
| Number of pages | 8 |
| Journal | Clinical Chemistry |
| Volume | 48 |
| Issue number | 10 |
| Publication status | Published - 1 Oct 2002 |
Fingerprint
Dive into the research topics of 'European concerted action on anticoagulation. Evaluation of a method for International Sensitivity Index calibration of two point-of-care prothrombin time (PT) monitoring systems (CoaguChek Mini and TAS PT-NC) with fresh plasmas based on whole-blood equivalent PT'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver