TY - JOUR
T1 - Evaluation of Aggregate and Silicone-oil Counts in Pre-filled Siliconized Syringes
T2 - An orthogonal study characterising the entire subvisible size range
AU - Shah, Maryam
AU - Rattray, Zahra
AU - Day, Katie
AU - Uddin, Shahid
AU - Curtis, Robin
AU - Van Der Walle, Christopher F
AU - Pluen, Alain
PY - 2017/3/15
Y1 - 2017/3/15
N2 - Characterisation of particulates in therapeutic monoclonal antibody (mAb) formulations is routinely extended to the sub-visible size-range (0.1–10 μm). Additionally, with the increased use of pre-filled syringes (PFS), particle differentiation is required between proteinaceous and non-proteinaceous particles such as silicone-oil droplets. Here, three orthogonal techniques: Raster Image Correlation Spectroscopy (RICS), Resonance Mass Measurements (RMM) and Micro-Flow Imaging (MFI), were evaluated with respect to their sub-visible particle measurement and characterisation capabilities. Particle formation in mAb PFS solutions was evaluated with increasing polysorbate-20 (PS-20) concentrations. All three techniques provided complementary but distinct information on protein aggregate and silicone-oil droplet presence. PS-20 limited the generation of mAb aggregates during agitation, while increasing the number of silicone-oil droplets (PS-20 concentration dependant). MFI and RMM revealed PS-20 lead to the formation of larger micron-sized droplets, with RICS revealing an increase in smaller sub-micron droplets. Subtle differences in data sets complicate the apparent correlation between silicone-oil sloughing and mAb aggregates’ generation. RICS (though the use of a specific dye) demonstrates an improved selectivity for mAb aggregates, a broader measurement size-range and smaller sample volume requirement. Thus, RICS is proposed to add value to the currently available particle measurement techniques and enable informed decisions during mAb formulation development.
AB - Characterisation of particulates in therapeutic monoclonal antibody (mAb) formulations is routinely extended to the sub-visible size-range (0.1–10 μm). Additionally, with the increased use of pre-filled syringes (PFS), particle differentiation is required between proteinaceous and non-proteinaceous particles such as silicone-oil droplets. Here, three orthogonal techniques: Raster Image Correlation Spectroscopy (RICS), Resonance Mass Measurements (RMM) and Micro-Flow Imaging (MFI), were evaluated with respect to their sub-visible particle measurement and characterisation capabilities. Particle formation in mAb PFS solutions was evaluated with increasing polysorbate-20 (PS-20) concentrations. All three techniques provided complementary but distinct information on protein aggregate and silicone-oil droplet presence. PS-20 limited the generation of mAb aggregates during agitation, while increasing the number of silicone-oil droplets (PS-20 concentration dependant). MFI and RMM revealed PS-20 lead to the formation of larger micron-sized droplets, with RICS revealing an increase in smaller sub-micron droplets. Subtle differences in data sets complicate the apparent correlation between silicone-oil sloughing and mAb aggregates’ generation. RICS (though the use of a specific dye) demonstrates an improved selectivity for mAb aggregates, a broader measurement size-range and smaller sample volume requirement. Thus, RICS is proposed to add value to the currently available particle measurement techniques and enable informed decisions during mAb formulation development.
U2 - 10.1016/j.ijpharm.2017.01.015
DO - 10.1016/j.ijpharm.2017.01.015
M3 - Article
SN - 0378-5173
VL - 519
SP - 58
EP - 66
JO - International Journal of Pharmaceutics
JF - International Journal of Pharmaceutics
IS - 1-2
ER -