Abstract
BACKGROUND AND PURPOSE: Target dose verification for VMAT treatments of head-and-neck (H&N) cancer using 3D in vivo EPID dosimetry is expected to be affected by daily anatomical changes. By including these anatomical changes through cone-beam CT (CBCT) information, the magnitude of this effect is investigated. MATERIALS AND METHODS: For 20 VMAT-treated H&N cancer patients, all plan-CTs (pCTs), 633 CBCTs and 1266 EPID movies were used to compare four dose distributions per fraction: treatment planning system (TPS) calculated dose and EPID reconstructed in vivo dose, both determined using the pCT and using the CBCT. D2, D50 and D98 of the planning target volume (PTV) were determined per dose distribution. RESULTS: When including daily anatomical information, D2, D50 and D98 of the PTV change on average by 0.0±0.4% according to TPS calculations; the standard deviation of the difference between EPID and TPS target dose changes from 2.5% (pCT) to 2.1% (CBCT). Small time trends are seen for both TPS and EPID dose distributions when using the pCT, which disappear when including CBCT information. CONCLUSIONS: Daily anatomical changes hardly influence the target dose distribution for H&N VMAT treatments according to TPS recalculations. Including CBCT information in EPID dose reconstructions slightly improves the agreement with TPS calculations.
Original language | English |
---|---|
Journal | Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology |
DOIs | |
Publication status | Published - 30 Jun 2015 |
Keywords
- Anatomical changes
- EPID dosimetry
- In vivo QA