Not all pacemakers are created equal: MRI conditional pacemaker and lead technology

Fozia Z. Ahmed, Gwilym M. Morris, Stuart Allen, Rajdeep Khattar, Mamas Mamas, Amir Zaidi

Research output: Contribution to journalArticlepeer-review

Abstract

MRI Conditional Pacemaker and Lead Technology Due to expanding clinical indications and an aging society there has been an increase in the use of implantable pacemakers. At the same time, due to increased diagnostic yield over other imaging modalities and the absence of ionizing radiation, there has been a surge in demand for magnetic resonance imaging (MRI) assessment, of both cardiac and noncardiac conditions. Patients with an implantable device have a 50-75% chance of having a clinical indication for MRI during the lifetime of their device. The presence of an implantable cardiac device has been seen as a relative contraindication to MRI assessment, limiting the prognostic and diagnostic utility of MRI in many patients with these devices. The introduction of MRI conditional pacemakers will enable more patients to undergo routine MRI assessment without risk of morbidity or device malfunction. This review gives a general overview of the principles and current evidence for the use of MRI conditional implantable cardiac devices. Furthermore, we appraise the differences between those pacemakers currently released to market. © 2013 Wiley Periodicals, Inc.
Original languageEnglish
Pages (from-to)1059-1065
Number of pages6
JournalJournal of cardiovascular electrophysiology
Volume24
Issue number9
DOIs
Publication statusPublished - Sept 2013

Keywords

  • implantable cardiac device
  • implantable cardioverter defibrillator
  • magnetic resonance imaging (MRI)
  • pacemaker

Fingerprint

Dive into the research topics of 'Not all pacemakers are created equal: MRI conditional pacemaker and lead technology'. Together they form a unique fingerprint.

Cite this