Performance of two blue light-emitting-diode dental light curing units with distance and irradiation-time

Adrian W. Bennett, David C. Watts

    Research output: Contribution to journalArticlepeer-review

    Abstract

    The performance of two blue light-emitting-diode (LED) curing units, in terms of their spectral output and irradiance and the depth of cure (d cure) produced in standard hybrid and modified composites was evaluated. Three composites of similar formulation but differing in their initiator concentrations and/or opacity were studied. dcure was measured using a calibrated digital needle-penetrometer, as a function of source-specimen distance and for irradiance periods of 10, 20 and 40 s. The results show that the cure-depth (dcure) values varied significantly with irradiance times and source-specimen distance for both LED and quartz tungsten halogen (QTH) sources.
    Original languageEnglish
    Pages (from-to)72-79
    Number of pages7
    JournalDental Materials
    Volume20
    Issue number1
    DOIs
    Publication statusPublished - Jan 2004

    Keywords

    • Composite shade
    • Dental material
    • Depth of cure
    • Exposure time
    • Light curing
    • Light-emitting-diode

    Fingerprint

    Dive into the research topics of 'Performance of two blue light-emitting-diode dental light curing units with distance and irradiation-time'. Together they form a unique fingerprint.