PRaVDA: The first solid-state system for proton computed tomography

M. Esposito, C. Waltham, J.T. Taylor, S. Manger, B. Phoenix, T. Price, G. Poludniowski, S. Green, P.M. Evans, P.P. Allport, S. Manolopulos, J. Nieto-Camero, J. Symons, N.M. Allinson

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose
Proton CT is widely recognised as a beneficial alternative to conventional X-ray CT for treatment planning in proton beam radiotherapy. A novel proton CT imaging system, based entirely on solid-state detector technology, is presented. Compared to conventional scintillator-based calorimeters, positional sensitive detectors allow for multiple protons to be tracked per read out cycle, leading to a potential reduction in proton CT scan time. Design and characterisation of its components are discussed. An early proton CT image obtained with a fully solid-state imaging system is shown and accuracy (as defined in Section IV) in Relative Stopping Power to water (RSP) quantified.
Method
A solid-state imaging system for proton CT, based on silicon strip detectors, has been developed by the PRaVDA collaboration. The system comprises a tracking system that infers individual proton trajectories through an imaging phantom, and a Range Telescope (RT) which records the corresponding residual energy (range) for each proton. A back-projection-then-filtering algorithm is used for CT reconstruction of an experimentally acquired proton CT scan.
Results
An initial experimental result for proton CT imaging with a fully solid-state system is shown for an imaging phantom, namely a 75 mm diameter PMMA sphere containing tissue substitute inserts, imaged with a passively-scattered 125 MeV beam. Accuracy in RSP is measured to be ⩽1.6% for all the inserts shown.
Conclusions
A fully solid-state imaging system for proton CT has been shown capable of imaging a phantom with protons and successfully improving RSP accuracy. These promising results, together with system the capability to cope with high proton fluences (2×108 protons/s), suggests that this research platform could improve current standards in treatment planning for proton beam radiotherapy.
Original languageUndefined
JournalPhysica Medica
DOIs
Publication statusPublished - 9 Nov 2018

Research Beacons, Institutes and Platforms

  • Manchester Cancer Research Centre

Cite this