Abstract
The internal optics of the recent models of the Shin-Nippon SRW-5000 autorefractor (also marketed as the Grand Seiko WV-500) have been modified by the manufacturer so that the infrared measurement ring has been replaced by pairs of horizontal and vertical infrared bars, on either side of fixation. The binocular, open field-of-view, allowing the accommodative state to be objectively monitored while a natural environment is viewed, has made the SRW-5000 a valuable tool in further understanding the nature of the oculomotor response. It is shown that the root-mean-square of model eye measures was least (0.017 ± 0.002D) when the separation of the horizontal measurement bars were averaged twice. The separation of the horizontal bars changes by 3.59 pixels/dioptre (r2 = 0.99), allowing continuous on-line analysis of the refractive state at up to 60 Hz temporal resolution to an accuracy of 3 mm. The pupil edge is not obscured in the diagonal axis by the measurement bars, unlike the ring of the original optics, so in the newer model pupil size can be measured simultaneously at the same rate with a resolution of
Original language | English |
---|---|
Pages (from-to) | 142-147 |
Number of pages | 5 |
Journal | Ophthalmic and Physiological Optics |
Volume | 24 |
Issue number | 2 |
DOIs | |
Publication status | Published - Mar 2004 |
Keywords
- Accommodation
- Autorefractor
- Continuous
- Dynamic
- Pupil