Ventricular oversensing in an implantable defibrillator during lead impedance testing

Matthew Luckie, Christopher Shepley, Lindsay Ford, Bernard Clarke, Amir Zaidi

    Research output: Contribution to journalArticlepeer-review

    Abstract

    Ventricular oversensing in an implantable defibrillator may be associated with inappropriate shocks. We report a case of oversensing during implantable cardioverter-defibrillator (ICD) implantation, which occurred recurrently during manual lead impedance testing. The oversensing is determined to be normal device function, explained by the programmed parameters, and algorithms determining the related aspects of device function. All rights reserved. © The Author 2010.