TY - JOUR
T1 - Ventricular oversensing in an implantable defibrillator during lead impedance testing
AU - Luckie, Matthew
AU - Shepley, Christopher
AU - Ford, Lindsay
AU - Clarke, Bernard
AU - Zaidi, Amir
PY - 2011/5
Y1 - 2011/5
N2 - Ventricular oversensing in an implantable defibrillator may be associated with inappropriate shocks. We report a case of oversensing during implantable cardioverter-defibrillator (ICD) implantation, which occurred recurrently during manual lead impedance testing. The oversensing is determined to be normal device function, explained by the programmed parameters, and algorithms determining the related aspects of device function. All rights reserved. © The Author 2010.
AB - Ventricular oversensing in an implantable defibrillator may be associated with inappropriate shocks. We report a case of oversensing during implantable cardioverter-defibrillator (ICD) implantation, which occurred recurrently during manual lead impedance testing. The oversensing is determined to be normal device function, explained by the programmed parameters, and algorithms determining the related aspects of device function. All rights reserved. © The Author 2010.